Target 96 Development

Gene
CD177

Uniprot
Q8N6Q3

Protein
CD177 antigen

See alternative names Polycythemia rubra vera protein 1,
NB1 glycoprotein,
Human neutrophil alloantigen 2a

Uniprot Function Description

In association with beta-2 integrin heterodimer ITGAM/CD11b and ITGB2/CD18, mediates activation of TNF-alpha primed neutrophils including degranulation and superoxide production (PubMed:21193407). In addition, by preventing beta-2 integrin internalization and attenuating chemokine signaling favors adhesion over migration (PubMed:28807980). Heterophilic interaction with PECAM1 on endothelial cells plays a role in neutrophil transendothelial migration in vitro (PubMed:17580308). However, appears to be dispensable for neutrophil recruitment caused by bacterial infection in vivo (PubMed:23461681). Acts as a receptor for the mature form of protease PRTN3 allowing its display at the cell surface of neutrophils (PubMed:17244676, PubMed:18462208). By displaying PRTN3 at the neutrophil cell surface, may play a role in enhancing endothelial cell junctional integrity and thus vascular integrity during neutrophil diapedesis (PubMed:23202369).

Sample type

Recommended sample types are EDTA plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsies, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary in different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Precision

Precision (repeatability) is calculated from linearized NPX values over LOD.

Within run precision Coefficient of Variation (CV)
9%
Between run precision Coefficient of Variation (CV)
10%

Analytical measuring range

The technical data reported below refers to the measured value in the in vitro validation assays run using known concentrations of recombinant antigen. Please note that these figures are for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples.

LOD (pg/mL)
0.480
LLOQ (pg/mL)
0.480
ULOQ (pg/mL)
7812
Hook (pg/mL)
7812
Range (logs)
4.2

Sensitivity plot

The calibrator curve shown below visualizes the analytical measuring range data based on in vitro measurement of recombinant antigen. Please note that this is shown for reference only and CANNOT be used to convert NPX values to absolute concentrations for proteins measured in plasma or serum samples. The vertical dotted lines represent LLOQ and ULOQ respectively, and the horizontal line indicates the LOD.

SensitivityPlot

Biomarker Validation Data

Additional validation data, as well as a more detailed description of how the Olink panels are quality controlled can be found in our Data Validation documents – go to Document download center

Search all biomarkers

Target 96 Development