Peptidyl-prolyl cis-trans isomerase FKBP4 (FKBP4)

Links to databases

Immunophilin protein with PPIase and co-chaperone activities. Component of steroid receptors heterocomplexes through interaction with heat-shock protein 90 (HSP90). May play a role in the intracellular trafficking of heterooligomeric forms of steroid hormone receptors between cytoplasm and nuclear compartments. The isomerase activity controls neuronal growth cones via regulation of TRPC1 channel opening. Acts also as a regulator of microtubule dynamics by inhibiting MAPT/TAU ability to promote microtubule assembly. May have a protective role against oxidative stress in mitochondria.

Sample type

Recommended sample types are human EDTA plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsis, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary in different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

NOTE: The calibrator curve below shows the performance of the assay with the estimated sensitivity and dynamic range parameters indicated. These curves are generated during the assay validation process using recombinant antigens. Please note that when analyzing biological samples the data generated will be given in the form of relative quantification (NPX values) and cannot be converted to absolute protein concentrations. For more info about NPX measurements, please visit our FAQ page.

Analytical Measuring Range

Please note: the technical data reported below refers to the measured value in the in vitro validation assays. Since the Metabolism panel uses human samples diluted 10-fold, a multiplication factor of 10 should be applied when comparing the addressable biological concentration to this validation data.

    • LOD (pg/mL)
    • 244
    • LLOQ (pg/mL)
    • 244
    • ULOQ (pg/mL)
    • 62500
    • Hook (pg/mL)
    • 62500
    • Range (logs)
    • 2.4

Calibrator curve for validation data (generated in multiplex)

Peptidyl-prolyl cis-trans isomerase FKBP4 (FKBP4)

Precision

    • Within run precision Coefficient of Variation (CV)
    • 9%
    • Between run precision Coefficient of Variation (CV)
    • 11%

Precision (repeatability) is calculated from linearized NPX values over LOD.

Biomarker Validation Data