Interleukin-1 beta (IL1B)

Links to databases

Potent proinflammatory cytokine. Initially discovered as the major endogenous pyrogen, induces prostaglandin synthesis, neutrophil influx and activation, T-cell activation and cytokine production, B-cell activation and antibody production, and fibroblast proliferation and collagen production. Promotes Th17 differentiation of T-cells. Synergizes with IL12/interleukin-12 to induce IFNG synthesis from T-helper 1 (Th1) cells (PubMed:10653850).

Sample type

Recommended sample types are EDTA plasma and serum. A range of additional sample types are compatible with the technology (PEA), including citrate plasma, heparin plasma, cerebrospinal fluid, (CSF), tissue and cell lysates, fine needle biopsis, microdialysis fluid, cell culture media, dried blood spots, synovial fluid, saliva, plaque extract and microvesicles. Please note that protein expression levels are expected to vary in different sample types. Certain assays are differentially affected by interfering substances such as hemolysate. Download any of our Data Validation documents or contact support@olink.com for more information.

Analytical Measuring Range

Please note: the technical data reported below refers to the measured value in the in vitro validation assays. The calibrator curves are generated during the assay validation process using recombinant antigens, with data presented as Normalized Protein eXpression (NPX) values plotted against protein concentration (in pg/mL). Please note that when analyzing biological samples the in vitro-generated curves presented here cannot be used to derive protein concentrations in pg/mL from the NPX values obtained. Olink Target 48 Cytokine delivers absolute protein concentrations directly using the calibrator system built into the system. For more information about NPX measurements, please view the FAQ.

    • LOD (pg/mL)
    • 0.095
    • LLOQ (pg/mL)
    • 0.19
    • ULOQ (pg/mL)
    • 3125
    • Hook (pg/mL)
    • 25000
    • Range (logs)
    • 4.8

Calibrator curve for validation data (generated in multiplex)

Interleukin-1 beta (IL1B)

Precision

    • Within run precision Coefficient of Variation (CV)
    • 5%
    • Between run precision Coefficient of Variation (CV)
    • 5%

Precision (repeatability) is calculated on quantified data within the limit of quantification: LOQ, between lower limit of quantification (LLOQ) and upper limit of quantification (ULOQ).

Download the complete panel data validation document for Olink Target 48 Cytokine> 